FAQ Software - Why am I getting an I/O Cancelled error in PowerSUITE with my PARSTAT potentiostat?
An I/O Cancelled error results when the booting process of the potentiostat is interrupted. The potentiostat must be fully booted before attempting to open, close or run an experiment. You can tell the instrument has fully booted when you see the “Booting Pstat XX%” message on the Instrument Bar change to “Idle”. You can then proceed with any operations. If you receive this error you must turn the instrument off and back on to clear it and then allow the instrument to finish the booting process before proceeding.
 